ICU呼吸机2000B2
10.4”TFT彩色液晶显示屏,操作简单便捷
Ventilation Mode:

VCV,PCV,SIMV,PSV,PSV SIMV,CPAP

Ventilation Function:

PEEP,SIGH,Standby,IP,IRV

warranty:

1 year

产品详情

2000B2呼吸机是一款采用气动、电子控制、微电脑技术的多功能呼吸机,采用进口元件,运行稳定可靠,10.4英寸彩色液晶屏显示各种测量和配置参数,非常适合呼吸治疗,适用于医院急救科、外科、妇科、儿科、ICU等科室。ICU科室使用需连接机上气源接口。

微信图片_20250313134138.png

显示模式               

高清10.4英寸TFT彩色LCD显示屏,操作简单便捷

 

通风模式          

VCV,PCV,SIMV,PSV,PSV SIMV,CPAP


通风功能         

PEEP,SIGH,待机,IP,IRV

 

Ventilation Parameters        

Tidal volume:20~2000ml

Rate:2~120bpm

SIMV rate:2~20bpm

I:E:4:1~1:8

Inspiratory trigger pressure:-10~20 cmH2O

Flow trigger sensitivity:1-20L/min

PEEP:0~25 cmH2O

Pressure control:3~60 cmH2O

Pressure limit:5~60 cmH2O

Pressure slope:1~10 gear

Flow rate setting:1~10 gear

Expiratory trigger sensitivity:10%~90%

Time control:1~10s

SIGH:1.5 times the inspiratory time(60-120 times adjustable)

Inspiratory platform:0~50%

Oxygen concentration:21%~99%

 

■Parameters For Ventilation Monitoring

VI tidal volume, VE tidal volume, Minute Ventilation volume,  IPPV rate, SIMV rate,  total respiratory rate, autonomous breathing rate, I/E, peak pressure of airway, average pressure, PEEP, Inspiratory trigger pressure, inspiratory platform, oxygen concentration, battery capacity , pressure - time waveform, flow rate - time waveform, flow-volume loop, pressure –volume loop

 

■Security Alarm System

Oxygen Concentration Alarm: Upper limit setting range:21%~99%,Low limit setting range:15%~80%

Airway Pressure Alarm: Upper limit setting range:1~60 cmH2O,Low limit setting range:0~50cmHO

Per-minute Ventilation Volume Alarm: Upper limit setting range:1~30L/min, Low limit setting range: 0~20L/min

Tidal Volume alarm: Upper limit setting range:10~2000ml,Low limit setting range   0~1800ml

Respiratory rate:Upper limit setting range 10~99bPm;Low limit setting range 0~50 bPm

Sustained High Pressure Alarm: It will give alarm when stress have consistently been higher than 25 cmH2O for 15s.

Low Battery Level Alarm: It will give alarm when battery level is less than 14±0.5V

Gas shortage alarm

Low pressure of Oxygen alarm

Low pressure of Air alarm

Power alarm                    

Intubation off alarm

Battery voltage alarm

 

■Power Supply

AC 100~240V,50/60Hz

Built-in Li Battery (Work at least for 4-6 hours with battery)